Appendix B
10
Appendix B
ODORATOR Correction Chart for Methane
This chart assumes the ODORATOR has been properly zeroed and allowed
to run for at least one minute.
Methane Correction Chart, 1.00 % Reference
ODORATOR %
Gas Reading
Actual % L.E.L. Actual % Gas
.10 1.4 0.08
.20 3.4 0.17
.30 5.0 0.25
.40 6.6 0.33
.50 8.0 0.40
.60 10.0 0.50
.70 12.4 0.62
.80 15.0 0.75
.90 17.4 0.87
1.00 Ref. 20.0 Ref. 1.00 Ref.
1.10 22.0 1.10
1.20 26.6 1.33
1.30 31.2 1.56
1.40 37.2 1.81
1.50 41.0 2.05