Heath-consultants Detecto Pak-Infrared (DP-I) Manual de usuario Pagina 22

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 35
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 21
102105-0 DP-IR MANUAL REV M
Page 22 of 35
Using the Bar Hole Probe
To use the Bar Hole Probe (102278-1; 20 in.) attach the probe to the DP-IR Filter cap using the
Quick Connector on the end of the Bar Hole Probe hose.
The Bar Hole Probe has two lters, the Hydrophobic Filter (102277-0) and the Particulate Filter
(102276-0), located in the Bar Hole Probe handle. Unscrew probe handle and ensure that both
lters are clean and dry. Replace if necessary.
Put the Bar Hole Probe handle back together, making sure handle is screwed tightly together and
O-ring makes a tight seal so no leaks will occur during survey. Also ensure that the sampling
holes at the end of the Bar Hole Probe are not blocked or plugged. The probe is now ready for
use.
Note: To ensure the Bar Hole Probe is sealed and assembled correctly, turn on the
instrument and block the holes at the end of the probe. The instrument should ow fault.
When conducting a Bar Hole survey, carefully watch the clear Bar Hole Probe during each
sample to ensure water is not being drawn up into the instrument. For more accurate sample
readings, regularly check the probe lters after survey samples to ensure the lters are clean and
dry.
Vista de pagina 21
1 2 ... 17 18 19 20 21 22 23 24 25 26 27 ... 34 35

Comentarios a estos manuales

Sin comentarios