1User Handbook
viiiUSER HANDBOOK3.3 BATTERY STATUS ... 3-33.4 TIME AND DATE ...
10-2USER HANDBOOK10.3 PRESSURE DISPLAYSWhen pressure test mode is selected from the menu, thenormal operating display is shown. See Fig. 10.1.Fig.
10-3PRESSURE MODE10.4 PREPARE INSTRUMENT (PRESSURE MODE)Zero the instrument before tubing is attached.To zero the instrument: Press and hold ZERO .C
10-4USER HANDBOOK10.5 PRESSURE BUTTON OPERATIONA summary of the button operation is detailed in Table10.1:LH(INVERT)RH(PUMP)CENTRE(RANGE)ACT
10-5PRESSURE MODE10.6.2 ZeroTo zero the instrument:(before tubing is attached) Press and hold the RH button .10.6.3 Menu / OffTo re-select
10-6USER HANDBOOK
11-1ALARMSALARMS11.1 GAS ALARMSThe following gas alarms are available according to gastype and are programmable according to application and/ or cu
11-2USER HANDBOOKNote: A Time Weighted Average (TWA) value is the meanaverage gas level over a specific period. The STELis 15 minutes and the LTE
11-3ALARMSIn the following examples, example 1 (Fig. 11.1) showsan instrument in confined space mode signalling a LOLOOxygen alarm. The a
11-4USER HANDBOOKExample 2 (Fig. 11.2) shows the instrument signalling aHIHI LEL alarm. Again the alarm warbles and all eight(8) LEDs ramp.
11-5ALARMSExample 3 (Fig. 11.3) shows an instrument in Leak TestMode with a Carbon Monoxide warning alarm. There is noaudible alarm and the LEDs d
ixCONTENTS5.8 CSM LOGGING ... 5-85.9 CSM CONFIDENCE SIGNAL ... 5-9BARHOLE TE
11-6USER HANDBOOKExample 4 (Fig. 11.4) shows an instrument in Leak TestMode with a HI LEL alarm. The audible alarm indicationis a high pitch tone
11-7ALARMSExample 5 (Fig. 11.5) shows an instrument in ConfinedSpace Mode with a LO O2 alarm. The audible alarmindication is a high
11-8USER HANDBOOKExamples 6 and 7 (Fig. 11.6 and Fig. 11.7) shows aninstrument in Confined Space Mode with Time WeightedAverage STEL and LTE
11-9ALARMSFig. 11.7 LTEL H2S Alarm11.5 ALARM TYPES11.5.1 Latching / Non LatchingEach alarm can be latching or non latching. Latching alarmsmu
11-10USER HANDBOOKThe following Table 11.1 shows the GMI default selections.Latching or non-latching options exist in all allowablealarms.Alarm
11-11ALARMSALARM TYPE LATCHINGYes / NoMUTEYes / NoAUDIBLEINDICATION VISUAL LEDINDICATIONDISPLAYLEL Warning N N / A N / A N / A Flash RangeLEL (HI)NY
11-12USER HANDBOOKThe following Table 11.2 shows the alarm options availablein the Leak Test and Confined Space Modes.LEAK TEST CSMFLAM CO H2S O2 FLAM
11-13ALARMS11.6 FAULT ALARMSRefer to Alarms Table 11.1 to identify the audible / visualindication for any of the following faults.11.6.1 Low
11-14USER HANDBOOKThe instrument can however still be used to detect andalarm on the other sensor(s) fitted.The faulty sensor will cause
11-15ALARMSIf fault remains, return instrument to an approvedService / Repair facility.alternating withFig. 11.9 Sensor Fault2) If a ZERO FAULT
xUSER HANDBOOK8.2 PURGE FEATURES ... 8-18.3 PURGE DISPLAYS ... 8-28
11-16USER HANDBOOKalternating withFig. 11.10 Sensor Fault11.6.4 Sample / Flow FaultIf a FLOW FAULT flag appears, alternating with mode asshown
11-17ALARMSFig. 11.11 Sample / Flow FaultCheck sample line, sample filter or probe for blockage, ifapplicable. Clear blockage then restart the pump.T
11-18USER HANDBOOK11.6.5 Calibration RequiredDuring warm-up, if the CALIBRATION REQUIRED flag isdisplayed and an audible alarm and Red LE
11-19ALARMS11.6.6 Service ExpiredDuring the warm up of the instrument, a check is done toverify if the service date has expired. If the instrument
11-20USER HANDBOOK
12-1OPERATOR MAINTENANCEOPERATOR MAINTENANCE12.1 CLEANINGCAUTION: Do not use polishes containing silicon orsolvent to clean the instrument
12-2USER HANDBOOK12.2.1 Probe Filter1) Hold the probe filter bulb then unscrew the probeconnector in a counter clockwise direction andremove th
12-3OPERATOR MAINTENANCEPROBE TUBEHYDROPHOBICFILTER(Part No. 67259)PROBE FILTER BULBFig. 12.2 Probe Filter Bulb Removal3) Hold the hydrophobi
12-4USER HANDBOOK4) Inspect the filter for contamination or damage.If either is evident, replace the filter.5) Fit a new Hydrophobic Probe F
12-5OPERATOR MAINTENANCE12.2.2 Chemical Filter (Accessory)The Chemical Filter assembly (Part No.67142) containswater absorbent silica gel
xiCONTENTSALARMS ... 11-111.1 GAS ALARMS ... 11-111.2 FL
12-6USER HANDBOOK2) Unscrew the chemical filter assembly in a counterclockwise direction, then remove the filter fromthe filter bulb as shown in
12-7OPERATOR MAINTENANCE5) The chemical housing filter disc must be replacedif it is damaged or contaminated.Using the flat end of a pencil or simila
12-8USER HANDBOOKFILTER HOUSINGADAPTORFILTER DISC(Part No. 67199)PENCIL(OR SIMILAR)Fig. 12.7 Adaptor Filter RemovalCAUTION The contents of th
12-9OPERATOR MAINTENANCE10) Attach chemical filter to filter bulb, in a clockwisedirection, then tighten to secure. See Fig. 12.4.Note: Care must
12-10USER HANDBOOK12.2.3 Cotton Filter (Accessory)The cotton filter assembly (Part No.67196) contains acotton filter to protect the inst
12-11OPERATOR MAINTENANCE2) Unscrew the cotton filter housing in a counterclockwise direction, then remove the housing fromthe filter bulb as s
12-12USER HANDBOOK7) Before attaching the probe connector to the cottonfilter housing, check the probe washer for signsof damage. If damaged, i
12-13OPERATOR MAINTENANCE12.3 BATTERY REPLACEMENTThe GT series instrument handle contains three batteriesthat provide the power required to operate
12-14USER HANDBOOK The Low Battery flag appears on thedisplay The instrument will not switch OnWhen the Low Battery flag appears on the disp
12-15OPERATOR MAINTENANCE12.3.1 Remove and Replace BatteriesWARNING 1: To prevent ignition of flammable orcombustible atmospheres, removebatt
xiiUSER HANDBOOKADDITIONAL INFORMATION ... 15-1Training ... 15-1Wor
12-16USER HANDBOOK1) Using a suitable screwdriver, unscrew in a counterclockwise direction and then remove the securingscrew from the battery cove
12-17OPERATOR MAINTENANCE3) Carefully lift then remove the three batteries fromthe compartment in the instrument handle, asshown in Fig. 12.12.N
12-18USER HANDBOOK4b) ALKALINE:Replace with three new LR 14 (Duracell, Energizeror Rayovac) alkaline batteries. Note polarity ofbatteries as sho
12-19OPERATOR MAINTENANCE6) Slide the battery cover over the instrument handleuntil the securing catch engages with theinstrument body.7) Check
12-20USER HANDBOOK A Commercial Battery Charger, available from mostelectrical outlets.Note: The batteries must be removed from the in-strument, as
12-21OPERATOR MAINTENANCEIf the instrument displays Charging Terminated, as shownin Fig. 12.16, then it has detected an excessive chargevol
12-22USER HANDBOOKStandard Instrument Charger (Universal Plug)(Part No. 67134)The instrument should be left overnight connected to thestandard charge
12-23OPERATOR MAINTENANCETo connect Standard Charger to the instrument:1) Lift dust cover from charger socket in rear face ofinstrument then connect
12-24USER HANDBOOKCharging Station(Part No. 67101)The charging station, shown in Fig. 12.19, is used tolocate the instrument securely whi
12-25OPERATOR MAINTENANCEPROBE LOCATIONSECURING STRAPPOWER ON INDICATORFig. 12.19 GT Series Charging Station
1-1INTRODUCTION1INTRODUCTIONThe GMI GT series instruments are designed to bemultifunction, multi-application gas detectors to suit all thene
12-26USER HANDBOOKTo install instrument in charging station:1) Check that instrument charging contacts, shownin Fig. 12.20, and the station
12-27OPERATOR MAINTENANCECHARGINGCONTACTSINSTRUMENT HANDLESTORAGE CLIPSECURINGSTRAPEXTENSION POLE LOCATIONIN INSTRUMENT HANDLEFig 12.21 Charging S
12-28USER HANDBOOK4) Insert probe in location in charging station, asshown in Fig. 12.19.5) If permanent 12V power supply is not install
12-29OPERATOR MAINTENANCE12V / 24V Vehicle Charger(Part No. 66206)The 12V / 24V Vehicle Charger, shown in Fig. 12.23,provides the option of
12-30USER HANDBOOK
13-113CALIBRATIONThe instrument has been calibrated for particular gases.Where any doubt exists the product should be returnedto GMI or an aut
13-2USER HANDBOOK13.1 CALIBRATION VALIDITYCalibration validity is the responsibility of the user. Undernormal operating conditions a 12 month p
14-114ACCESSORIESAccessories available for theGT series instruments areas follows:Std. AccessoriesPart Number Description67108 Carrying Case6
14-2USER HANDBOOK67120 Protective Rubber Boot67134 Standard Instrument Charger(Universal Plug)66206 12V/24V Vehicle InstrumentCharger67101 Charging
15-115ADDITIONAL INFORMATIONTrainingTraining courses are available on all GMI products. ContactGMI Marketing Department for further details:Tel: +44
1-2USER HANDBOOKIn the following procedure, the five instrument buttonsshown in Fig 1.2 are referred to as detailed below:UpLHRHCentreDownFig.
15-2USER HANDBOOK
A-1TYPICAL OPERATING PARAMETERSTYPICAL OPERATINGPARAMETERSTypical operating parameters are as follows:Gas Range ResolutionRangePPM 0 to 10000 1
A-2USER HANDBOOKDimensions (excl. probe)290mm (11.4) x 95mm (3.7) x 55mm (2.2)Weight (incl. probe & batteries)0.78kg (1.7lbs.)Temperature Limi
IINDEXIndexSymbols(LEL) ALARMS 11-1(O2) ALARMS 11-1AAcknowledging 11-9ADVICE, DISPOSAL iALARM TYPES 11-9ALARMS 11-1ALARMS, (LEL) 11-1ALARMS,
IIUSER HANDBOOKCO BUTTON OPERATION7-6CO DISPLAYS 7-3CO FEATURES 7-1CO LOGGING 7-8CO MENU 7-2CO RANGES 7-1CO) ALARMS 11-1Confined Space 2-2CONFI
IIIINDEXFFAULT ALARMS 11-13Fault, Sample 11-16Fault, Sensor 11-14Fault, Zero 11-13FEATURES, BARHOLE 6-1FEATURES, CO 7-1FEATURES, CSM 5-1FEATURE
IVUSER HANDBOOKMMAINTENANCE, OPERA-TOR 12-1MANUAL CALIBRATIONSYSTEM 13-1, 14-1MENU, CO 7-2MENU, MODE 3-17MODE MENU 3-17MODES OF OPERATION 2-2MOD
VINDEXRRANGES, BARHOLE 6-1RANGES, CO 7-1RANGES, CSM 5-1RANGES, INSTRUMENT 1-3RANGES, PRESSURE 10-1RANGES, PURGE 8-1RANGES, SNIFFER 9-1Rating, P
VIUSER HANDBOOKZZero Fault 11-13
1-3INTRODUCTIONThe bottom line of the instrument display may indicatebutton press options. For example, to select Yes fromthe follo
GMI Head Office:Inchinnan Business Park,Renfrew,PA4 9RG,Scotland, U.K.Telephone +44 (0)141 812 3211Fax +44 (0)141 812 7820e-mail: [email protected]
1-4USER HANDBOOK
2-1GENERAL FEATURESGENERAL FEATURES An integral pump draws the required sample overall of the sensors The pump flow is monitored by means of a
2USER HANDBOOKIssue 230/05/06Part Number: 67112GMI welcomes comments on all our publications.Your comments can be of great value in help
2-2USER HANDBOOKBy simple selection of the application you need, theappropriate gas(es) are shown on the display and onlythe necessary
2-3GENERAL FEATURESPurgeThis mode is used in gas and air purging applications.Refer to Chapter 8 for Purge mode operation.SnifferThis mode is use
2-4USER HANDBOOK
3-1OPERATING PROCEDUREOPERATING PROCEDURECheck the following: The instrument is clean and in good condition. The batteries are in good condition,
3-2USER HANDBOOK Each time you use the instrument, carry out the followingprocedure:3.1 SWITCH THE INSTRUMENT ONTo switch the instrument ON in f
3-3OPERATING PROCEDURE3.2 INSTRUMENT IDENTIFICATIONDuring the warm-up cycle, the instrument display identifiesthe model, serial number, software
3-4USER HANDBOOK3.4 TIME AND DATEThe time and date from the instruments built-in clock isdisplayed on the screen during warm-up, as shown
3-5OPERATING PROCEDURE2. Cal Due Date and overdue date messages aredisplayed.Calibration due date is displayed, as shown in Fig.3.5.Fig. 3.5 Cali
3-6USER HANDBOOK3. Cal Due Date message is displayed with useracknowledge if overdue (Default).Calibration due date is displayed as shown in
3-7OPERATING PROCEDURETo switch OFF: Press and hold NO . The screen, shown inFig. 3.9, is displayed.Fig. 3.9 Switch OFF Press and hold bot
iCOPYRIGHTCOPYRIGHTThis User Handbook is copyright of Gas Measurement Instruments Ltd (GMI) andthe information contained within, is for use only with
3-8USER HANDBOOKIf overdue but within the extended period, thescreen, shown in Fig. 3.11, is displayed.Fig. 3.11 Calibration OverdueThe
3-9OPERATING PROCEDUREFig. 3.12 Switch OFF Press and hold both the LH button and the RHbutton simultaneously to proceed with theswitch-off
3-10USER HANDBOOKFig. 3.14 Switch OFFTo proceed with the switch-off sequence. Press and hold both the LH button and the RHbutton simultaneo
3-11OPERATING PROCEDURE2. Service Due Date and overdue date messagesare displayed, as shown in Fig. 3.15. Fig. 3.15 Service Due DateThe screen,
3-12USER HANDBOOK3. Service Due Date message is displayed withuser acknowledge if overdue.Service due date is displayed, as shown in
3-13OPERATING PROCEDURETo switch OFF: Press and hold NO . The screen, shown inFig. 3.19, is displayed.Fig. 3.19 Switch OFF Press and hold
3-14USER HANDBOOKIf overdue but within the extended period, thescreen, shown in Fig. 3.21, is displayed.Fig. 3.21 Service OverdueThe user
3-15OPERATING PROCEDUREFig. 3.22 Switch OFF Press and hold both the LH button and the RHbutton simultaneously to proceed with theswitch-of
3-16USER HANDBOOKThe screen, shown in Fig. 3.24, is displayed whenthe Service date has expired. i.e. overdue.Fig. 3.24 Switch OFFTo proceed with
3-17OPERATING PROCEDURE3.7 SWITCH THE INSTRUMENT OFF OR RE-ENTER THE MODE MENUTo initiate the shut down sequence: Press and hold both the LH and R
iiUSER HANDBOOKSAFETY The instrument must be regularly serviced and calibrated by fully trainedpersonnel in a safe area. Batteries: Alkaline or *Re
3-18USER HANDBOOKalarms are active, the user must press and holdboth the LH and RH buttons simultaneouslyfor an additional five (5) seconds
3-19OPERATING PROCEDURETo highlight the required option: Press the UP or DOWN buttons, then . .To select the highlighted option: Press and hold O
3-20USER HANDBOOK
4-1LEAK TEST MODELEAK TEST MODEThis mode is for the technician to investigate odour orleak complaints and to pinpoint the leak.4.1 LEAK TE
4-2USER HANDBOOK Manual Range selection between Flam and Flamplus CO (Dual Display) Threshold adjustment Display Invert Automatic datalogging4.3
4-3LEAK TEST MODEFig. 4.2 Normal Display ExampleNote 1: Any faulty sensor will continue to display a spanner(wrench) symbol alternating with the
4-4USER HANDBOOKWhen the PPM autoranges to LEL, the display changes tothe example shown in Fig. 4.4.Fig. 4.4 LEL DisplayIf displaying the maximum rea
4-5LEAK TEST MODEIf a CO sensor is included in your instrument and the COconcentration rises above a pre-set level, the display willchange together
4-6USER HANDBOOKFig. 4.7 Alarm AcknowledgeWhen CO levels are below alarm concentration, the Centrebutton enables return to flammable only. This bu
4-7LEAK TEST MODE4.5 DESCRIPTION OF LEAK TEST BUTTONOPERATION4.5.1 Ticker (Geiger) ON / OFFTo enable and disable the audible Ticker (Geiger):(in th
iiiCOPYRIGHTAREAS OF USEExposure to certain chemicals can result in a loss of sensitivity of the flammablesensor. Where such environments are known or
4-8USER HANDBOOKFig. 4.8 Ticker (Geiger) Threshold4.5.3 Range SelectionThe instrument initially selects the PPM range whichautoranges
4-9LEAK TEST MODE4.5.5 Max / LiveTo show the Maximum readings:(since the mode was selected) Press and hold the LH button .To return display to
4-10USER HANDBOOK4.5.7 PumpTo switch the pump ON and OFF: Press the RH button 4.5.8 ZeroTo zero the Flammable PPM range when displayed: Press
4-11LEAK TEST MODEWhen the menu appears on the screen, releasethe buttons, otherwise the instrument will proceedinto the switch OFF process
4-12USER HANDBOOKFig. 4.9 PPM Level / LED Sequence4.7 LEAK TEST ALARMSRefer to Chapter 11.4.8 LEAK TEST LOGGINGAutomatic datalogging is active f
5-1CONFINED SPACE MODECONFINED SPACE MODEThis mode is used for confined space pre-entry testingand for personal monitoring in areas such as
5-2USER HANDBOOK Manual Logging is also available. This will log allgas concentrations at that instant. Time weighted averaging for Alarms
5-3CONFINED SPACE MODEWhen an instrument has an alarm active, the bottom lineof the display changes. A sample fault will change bothtop and bo
5-4USER HANDBOOKWhen viewing Min in confined space mode, Confined Minidentifier is in the top of the display, with the option ofselecting L
5-5CONFINED SPACE MODE5.5 DESCRIPTION OF CSM BUTTON OPERATION5.5.1 LightsTo switch ON the backlight: Press and hold the Centre button .To switch
ivUSER HANDBOOK
5-6USER HANDBOOKNote 2: If no button is pressed, after 30 seconds thedisplay reverts to live reading.5.5.3 Alarm AcknowledgeHIHI, LOLO for O
5-7CONFINED SPACE MODE5.5.4 Flow FaultBefore clearing a flow fault alarm the instrument must bechecked for water ingress or blockage.To clear a
5-8USER HANDBOOKNote: In Confined Space Mode, to prevent inadvertentlyswitching instrument OFF or changing mode whilealarms are active, the user
5-9CONFINED SPACE MODE5.9 CSM CONFIDENCE SIGNALDuring normal operation, the instrument sounds aconfidence beep and illuminates the bottom
5-10USER HANDBOOK
6-1BARHOLE TESTING MODE6BARHOLE TESTING MODEThis mode is used to locate underground leaks and onstart-up will be ranged to 0-100% LEL
6-2USER HANDBOOK6.3 BARHOLE DISPLAYSDuring the warm-up period, all applicable sensors will bechecked and the display will indicate any sensor
6-3BARHOLE TESTING MODETo highlight the required option: Press the UP or DOWN buttons, then . .To select the highlighted option: Press and hold O
6-4USER HANDBOOKTo change the sample time: Press the UP or DOWN buttons.Sample time range = 10 to 300 seconds.Holding the button will perform
6-5BARHOLE TESTING MODETo stop purge: Press and hold OK .The first barhole screen is then displayed:Fig. 6.5 Barhole TimerUp to six barhole
vREVISION RECORDDate Issue Description Of Change07/09/2005 1 New Handbook30/05/2006 2 To incorporate effect ofCR 4223, CR 4229, CR4279 and various
6-6USER HANDBOOKFig. 6.6 Barhole Numbering Between Barhole tests there is a mandatory Purgemode to ensure that any gas in the instrument
6-7BARHOLE TESTING MODE6.3.2 Non Timed ModeIf NON TIMED is selected or has been pre-selected inconfiguration then the instrument per
6-8USER HANDBOOKUp to six barhole readings can be stored. These areidentified as Barhole 1 to Barhole 6.After the first reading is stored
6-9BARHOLE TESTING MODEFig. 6.10 View Barhole ResultsTo return the display to normal operation: Press and hold LIVE .If a Flow Fault or Bead Fa
6-10USER HANDBOOK6.5 BARHOLE OPERATIONA summary of the button operation is detailed in thefollowing table:LH(INVERT)RH(PUMP)CENTRE(RANGE)AC
6-11BARHOLE TESTING MODETo switch ON the flashlight: Press and hold the Centre button again.The backlight and flashlight are both timed to switch OF
6-12USER HANDBOOK.6.6.5 Menu / OffTo re-select the Mode Menu: Press and hold both the LH button and the RHbutton .While both buttons are depre
7-1CO (CARBON MONOXIDE) MODECO (CARBON MONOXIDE) MODEThe CO mode is used to check the interior of premisesand appliances for CO leakage.7
7-2USER HANDBOOK7.3 CO MENUOn selection of CO mode, all three of the CO sub-modesshown in Fig. 7.1 can be available to the user.(See note 3)
7-3CO (CARBON MONOXIDE) MODE7.4 CO DISPLAYSDuring the warm-up period, all applicable sensors will bechecked and the display will indicate any
viUSER HANDBOOK
7-4USER HANDBOOK7.4.1 CO DirectIn this option, normal atmospheric air is checked for COcontent.The typical display for this mode is shown in Fig. 7.3
7-5CO (CARBON MONOXIDE) MODE7.4.3 Air-Free COAir-free CO is the CO reading modified by the O2 reading,therefore this measurement will only be
7-6USER HANDBOOK7.5 CO BUTTON OPERATIONA summary of the button operation is detailed in Table 7.1:LH(INVERT)RH(PUMP)CENTRE(RANGE)ACTIONPRESSINVERTDI
7-7CO (CARBON MONOXIDE) MODE7.6.2 Manual ZeroOnly available in differential CO mode.To zero differential CO: Press and hold ZERO .7.6.3 Manua
7-8USER HANDBOOK7.6.5 Menu / OffTo re-select the Mode Menu: Press and hold both the LH button and the RHbutton .While both buttons are depresse
8-1PURGE MODEPURGE MODEThe purge mode is used in gas and air purging applications.8.1 PURGE RANGESPurge mode will have the following ranges availabl
8-2USER HANDBOOK8.3 PURGE DISPLAYSDuring the warm-up period, all applicable sensors will bechecked and the display will indicate any sensor f
8-3PURGE MODETo display both % Gas and % O2:(as shown in Fig. 8.3) Press the Centre button Fig. 8.3 Gas and Oxygen Display8.4 PURGE BUTTON OP
8-4USER HANDBOOK8.5 DESCRIPTION OF PURGE BUTTONOPERATION8.5.1 LightsTo switch ON the backlight: Press and hold the Centre button To switch ON the f
8-5PURGE MODE8.5.4 Menu / OffTo re-select the Mode Menu: Press and hold both the LH button and the RHbutton .When the menu appears on th
viiCONTENTSCONTENTSCOPYRIGHT ... iLIABILITY ...
8-6USER HANDBOOK
9-1SNIFFER MODESNIFFER MODEThis mode is used to find small fitting leaks. Fast detectionrates are achieved using a semiconductor sensor in thepro
9-2USER HANDBOOKalternating withFig. 9.1 Sensor CheckTo zero the reading (in fresh air) Press and hold ZERO .Note: Pump must be switched ON to zer
9-3SNIFFER MODE9.4 SNIFFER BUTTON OPERATIONA summary of the button operation is detailed in Table 9.1:LH(INVERT)RH(PUMP)CENTRE(RANGE)ACTIONPRESSINVE
9-4USER HANDBOOK9.5.2 PumpTo switch pump ON and OFF: Press the RH button 9.5.3 ZeroTo zero the ppm range Press and hold the RH button .Note:
9-5SNIFFER MODE1. Press and hold - to enable visual only.2. Press and hold again - to enable both audible andvisual.9.5.6 Ticker (Geiger) AdjustT
9-6USER HANDBOOK9.6 SNIFFER TICKER (GEIGER) INDICATIONThe Ticker (Geiger) range at start up is 0-1000 ppm.For any subsequent PPM concentration, the T
9-7SNIFFER MODEFig. 9.3 Ticker (Geiger) LED Illumination
9-8USER HANDBOOK
10-1PRESSURE MODEPRESSURE MODEIn this mode, the instrument can be used as a manometerto measure appliance and regulator pressure and to checkthe syst
Comentarios a estos manuales