Heath-consultants GT Series Manual de usuario Pagina 131

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 170
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 130
12-7
OPERATOR MAINTENANCE
5) The chemical housing filter disc must be replaced
if it is damaged or contaminated.
Using the flat end of a pencil or similar, push the
filter from the sample side of the housing as shown
in Fig.12.6, then remove and discard the filter.
Insert new filter disc (Part No. 67138) into chemical
housing then, using the flat end of a pencil or similar,
carefully push filter into housing until correctly
seated.
FILTER DISC
(Part No. 67138)
FILTER HOUSING
Fig. 12.6 Housing Filter Removal
6) The housing adaptor filter must also be replaced if
it is damaged or contaminated.
Using the flat end of a pencil or similar, push the
filter from the open end of the adaptor as shown
in Fig.12.7, then remove and discard the filter.
Insert new filter disc (Part No. 67199) into adaptor
recess until correctly seated.
Vista de pagina 130
1 2 ... 126 127 128 129 130 131 132 133 134 135 136 ... 169 170

Comentarios a estos manuales

Sin comentarios