Heath-consultants GT Series Manual de usuario Pagina 139

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 170
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 138
12-15
OPERATOR MAINTENANCE
12.3.1 Remove and Replace Batteries
WARNING 1: To prevent ignition of flammable or
combustible atmospheres, remove
batteries before servicing.
WARNING 2: To prevent ignition of flammable or
combustible atmospheres, read,
understand and adhere to the
manufacturers live maintenance
procedures.
WARNING 3: To reduce the risk of ignition of a
flammable or explosive
atmosphere, batteries must be
changed only in a location known to
be non-hazardous.
WARNING 4: To reduce the risk of explosion, do
not mix old batteries with used
batteries or mix batteries from
different manufacturers.
WARNING 5: Never attempt to recharge non
rechargeable cells.
CAUTION 1: Not for use in oxygen enriched
atmospheres.
CAUTION 2: Replace batteries only with
approved batteries:
NiMH  GP350
Alkaline Duracell, Energizer or
Rayovac.