Heath-consultants Gas Delivery Unit (GDUnet) Manual de usuario Pagina 98

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 109
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 97
6-4
USER HANDBOOK
3. Select the drop down menu arrow, illustrated in Fig. 6-5, to
net as the calibration station and gas delivery
unit suitable for all GMI instrument types:
Fig. 6-5 Select Rig
4. 
be loaded (Nos. 2 to 8), together with ‘Zero Air’, always listed
as No. 1. ‘Zero Air is used for zeroing the instrument’s ranges
and cannot be edited.
The Gasurveyor 526 features four gas ranges, calibrated for
Methane as follows:
0 to 1000 PPM Flammable
0 to 100% LEL
 
0 to 1000ppm Carbon Monoxide
Remember that template gases / values entered must
be compatible with the corresponding gases / values on
each gas cylinder label.
5. To s pec i fy g a s No. 2 ( 0 to 1 0 00 PP M Fla m mab l e), s ele c t
to
display gases drop down list, as illustrated in Fig. 6-6.
Vista de pagina 97
1 2 ... 93 94 95 96 97 98 99 100 101 102 103 ... 108 109

Comentarios a estos manuales

Sin comentarios