Pagina 1 - HDR Series

Heath Data Recorder HDR Series User’s Manual & Field Installation Guide Heath Consultants Incorporated Houston, TX 713-844-1300 Fax:

Pagina 2

10 Internal Modem The HDR may be equipped with an internal modem and phone line surge suppressor. The modem is pre-configured and pre-wired from the

Pagina 3 - Table of Contents

11 HDR Host Operating Software HDR Host is the companion software interface to the HDR instrument. The application program provides for configuration,

Pagina 4

12 Accept the agreement. Then click “Next” to continue. Click “Next” to continue or change the default location of the HDR Host data files.

Pagina 5 - HDR Features

13 Click “Next” to continue or change the default location of the data log files. Click “Next” to continue or change the default location of the HD

Pagina 6 - HDR PCB LAYOUT

14 Select additional icons. Click “Next” to continue. Click “Next” to continue installation.

Pagina 7 - Mounting & Installation

15 Click “Finish” to complete the installation. Select the box to view the “README.TXT” file. Quick Start Guide 1. Run “HDR Host Setup 4.02.exe”

Pagina 8

16 HDR Configuration- Detailed User Friendly Names (Allows custom channel names of active channels- when required) 1. Go to the Configure Column/ “Us

Pagina 9 - GND (Pin 5)

17 Pulse outputs 1. The HDR has two pulse outputs that can be independently configured for Alarm closure (for external visual or audible alert or to

Pagina 10 - (with “Call out upon Alarm”)

18 Alarms Configuration (set up to 16 alarms as necessary): 1. Click the box next to the alarm to activate 2. Type the “Name” of the alarm to confi

Pagina 11 - HDR Host Operating Software

19 Configure LCD sequence 1. Go to the configure column/ “LCD Display Sequence”. Select the items to be viewed on the HDR LCD screen. Leave any items

Pagina 12

2 HDR Series HDR User’s Manual & Field Installation Guide Proprietary Notice The contents of this manual are proprietary to Heath Consultant

Pagina 13

20 2. Watch log: Records 1 pressure channel (High resolution log selectable from 1 second data, for 9 hours data storage, up to 16 second averaged d

Pagina 14

21 Set Access Control (This sets the security passwords of the instrument) 1. In HDR Host, Go to the Configure Column/ “Set Access Control” in the i

Pagina 15 - Quick Start Guide

22 Auto Configure (Allows saving and sending the Configuration, Calibration, Transducer Poly & site info) 1. To Save this file a. Go to "Co

Pagina 16 - HDR Configuration- Detailed

23 Modbus Implementation 1. Modbus can be used as long as it is ACSII (Modbus RTU is not currently active) Packet framing: Data bits=7 / Parity= Even

Pagina 17 - Pulse outputs

24 Configuring for use with an IP modem 1. Select “Host Setup – Configure Host”. 2. Click the “Network” Port Selection and then click “OK”.

Pagina 18

25 3. Select “Comm – Master Directory”. 4. Click on the “New IP Addr” button.

Pagina 19 - Configure LCD sequence

26 5. Enter the IP address, Port Number and choose a Name/Location for the installation. In the COM Selection drop-down box, click on “Select 4 – TCP

Pagina 20 - Set Real Time Clock

27 8. Double-click the IP modem’s site name or single click the site name and then click “OK”. Editing is also available with this screen if needed.

Pagina 21

28 10. Once connected, the Host screen will indicate the connected IP address in yellow at the bottom of the window. 11. Select “Security – L

Pagina 22

29 12. To illustrate remote IP modem communication, select “Display – Display 1 Live Readings”. 13. The Live Readings window will appear aft

Pagina 23 - Modbus Implementation

3 Table of Contents HDR I.S. Series (Purpose of this manual) ...

Pagina 24

30 14. Remote HDR IP modem communications are available for use using the HDR Host. When finished, exit remote IP modem communications by selectin

Pagina 25

31 HDR 325 Profiler Memory Capacity Table Example: 10 min recording of 4 fields yields 350 days of Profiler data stored in the HDR The Profiler is sel

Pagina 26

32 HDR Firmware Update Procedure (This update will not delete any Cal, CFG or saved data) To get the firmware file, contact Heath Consultants Inc. tec

Pagina 27

33 #3 Same Enter YOUR name here Select this to save Cal & CFG #4 #5 #6 Select this to default Cal & CFG

Pagina 28

34 Active during update Programming firmware Successful Exit to normal operation #8 #7 Watch & wait #9 Select Cold Start to clear all logs

Pagina 29

35 New Firmware should look like this in “Display/ View firmware”. #10 Reset Date & Time Reset Date & Time #10

Pagina 30

36 HDR Specifications Functionality Profiler Data Recording: Circular log: Large memory, recording intervals include: 1 min., 5 min., 10 min., 15 min

Pagina 31

37 HDR Maintenance and Troubleshooting The following may serve as a quick troubleshooting guide if you encounter problems in operating the HDR Heath D

Pagina 32

38 When calling the factory or your area representative for further troubleshooting, please have the following information ready: Customer Informat

Pagina 33

39 Maintenance Procedures Note: During the calibration process the last read pressure and gas temp will be used (LCD shows “- - CAL- -“) to calculate

Pagina 34 - Exit to normal operation

4 External Power Connector Heath Data Recorder (HDR) Overview The HDR is a microprocessor based, self-contained system designed for the purpose of pro

Pagina 35 - Reset Date & Time

40 Back-up Battery Replacement Procedure This procedure describes how to replace the lithium back-up battery BT1 (HPN 77R65-8001) and the profiler ba

Pagina 36 - HDR Specifications

41 7. Install the retainer clip over the lithium battery and verify both latches seat properly onto the holder.

Pagina 37

42 Spare Parts & Accessories Pressure Transducer The HDR may be equipped with a wide variety of ranges, most of which are available in either gaug

Pagina 38

43 Service Information: Warranties and Warranty Repair The HDR is warranted to be free from defects in material and workmanship for four (4) years

Pagina 39 - Maintenance Procedures

44 Configuration Worksheet Complete this worksheet prior to leaving the meter shop to facilitate field configuration and documentation. Always reme

Pagina 40

45 Technical Support 1-800-HEATHUS (1-800-432-8487) Heath Consultants Incorporated Factory Service Center 9030 Monroe Road Houston, Texas 77061

Pagina 41

5 HDR Features • Handles up to three pressures (3) and two (2) temperatures (plus ambient). • Intrinsically safe: Class I Div I Group D (without

Pagina 42 - 77R61-9580

6 HDR PCB LAYOUT CTSRTSCDTXMRXMGNDGNDPULSE INPUTGNDTXDRXDPULSE OUT BGNDPULSE OUT AGNDP3 +EP3 +SP3 -EP3 -SGNDP2 +EP2 +SP2 -EP2 -SGNDP1 +EP1 +SP1 -EP1 -

Pagina 43 - Obtaining Service

7 Mounting & Installation Housing Configuration The HDR is available in many configurations. The maximum number of integral sensors is as follows:

Pagina 44 - Configuration Worksheet

8 Battery Pack and Back-up Batteries The HDR units utilize (1) 1.5-volt, encapsulated “D” cell alkaline battery as its primary power source. The pack

Pagina 45 - 1-800-HEATHUS

9 External Modem The HDR is equipped with two RS232 communication ports. The Direct Port is dedicated to local direct PC communication at 19200 baud