HDR Maintenance and Troubleshooting
The following may serve as a quick troubleshooting guide if you encounter problems in operating the HDR Heath Data Recorder.
If any of these hints fail, call factory for assistance:
Symptom
Diagnosis
Will not log on
Verify that proper com port is selected in HDR Host/ “Host Setup”. (If a Serial to USB
adapter is used, find out the proper com port selection in the Windows “Device Manager”)
If error “Com not ready to send” check to see that only 1 session of HDR Host is running
and that no other program has the port open. After correcting problem, it may be necessary
to reboot the computer.
Make sure that “Default Sandia ID” in “Host Setup” is set to 63.
Check Com Cable
Check battery voltage
Will not scroll
Check battery voltage
Test scroll switch with ohmmeter
Restart instrument. Remove main power connector and then remove the lithium jumper on the
PCB. Wait 5 minutes, reconnect the main power first (with good battery), then re-connect the
lithium jumper (No data will be lost after restart).
Display is blank
Check Main battery voltage (if good, skip to step 4).
Check lithium voltage
Check lithium jumper is installed
Restart instrument. Remove main power connector, and then remove the lithium jumper (JP1) on
the PCB. Wait 5 minutes, re-connect the main power first (with good battery), then re-connect
the lithium jumper (JP1)(No data will be lost after restart).
Not responding
to applied
pressure
Check connections of transducer wires.
Send the original Calibration file” into the unit and verify calibration, Config column/
Auto Config”/ “Restore Instrument” Example filename: 01234567. PCF
Pressure not
correct in Profiler
(LCD is correct)
Reset profiler to update the Profiler header. (“Configure” Column/ “Reset Instrument/ Select
“Reset Profiler”/ Click “OK”.). At the “Confirm” screen, click “Yes”. Warning: This will erase
all Profiler data. If data is needed, download Profiler data first.
Not responding
to temperature
(Possible
“Fault” display)
Check connections of temperature probe wires.
Send original calibration file into the unit and verify calibration.
Replace with a known good temperature probe, retest & recalibrate.
Instrument shows
“Fault” (T-probe
or power cycle)
Reset the “Fault” by going to the “Configure” Column/ “Reset Instrument”. Select “Reset Alarm
and Fault logs”.
Instrument is
locked up or
not functioning
correctly
Reload instrument firmware. Remove all power (Main batt & Lithium jumper-JP1). Wait 2 mins.
Hold scroll button, then connect the main battery first, then connect lithium jumper. The LCD
should show - - Lodr- -. Follow the procedure on page 32-35 selecting “Firmware (W default Cfg
& Cal)” and “Cold Start” to ensure all memory is cleared. Then send config file: Config/ “Auto-
config”/ “Restore Instrument”/ select instrument filename/ Restore All.
Error
downloading
Profiler file:
“Invalid Host
File.
Rename, move or delete the host profiler file and then re-download the Profiler data. This allows
a new file to be created.