PROBLEM SOLUTION
“Light” alarm occurs: Clean lenses. If alarm
continues, re-align optics.
“Warm-up” alarm occurs: Turn light off until system
warms up.
“Failure” alarm occurs: Exit sample mode while
communicating through
RS-232 port.
“Batt Low” alarm occurs: If it occurs while vehicle
engine is off, start engine.
If it occurs while engine
is running, have your
battery and charging
system checked by a
mechanic.
Water has condensed on the The receiver has obtained
inside of the receiver lens: a leak, return the OMD
to the factory for repair.
Water has condensed on the The light assembly has
inside of the light assembly lens: obtained a leak, return the
OMD to the factory for
repair.
TROUBLESHOOTING