SYSTEM SPECIFICATIONS
OMD
TM
System Specifications
Configuration: Double - ended
Sensitivity: 1 PPM / meter CH4
Measurement Range: 1 to 200 PPM
Accuracy: ± 10% (1 to 100 PPM range)
± 20% (100 to 200 PPM range)
Above 200 PPM = Not relatively
accurate
Display Ranges: 10, 30 and 90 PPM
Self - test: During boot – up
Calibration/Test: Via operator, self - contained
Calibration: Via RS-232 through software
Base - Line Compensation: Via RS-232 through software
Display: Back-lit 2" x 6" graphics LCD,
adjustable contrast
Operator Interface: Sealed membrane switch overlay
Operator Alarms:
Signal: High pitch audible with
adjustable set point, pitch
increases with concentration.
Adjustable volume.
Error: Low pitch audible with
adjustable set point
Error alarms
provided for: Warm-up, Low Light,
Communication Failure &
Battery low
System Power: 72 Watts @ 12 VDC
System Voltage: 12 - 16 VDC
NOTE: To avoid damage to the OMD power box, the fuse
must be removed prior to charging or jumping the battery
in the survey vehicle!