Heath-consultants Hi Flow Sampler Manual de usuario Pagina 44

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 62
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 43
Maintenance Hi Flow
Sampler
44 Instruction 0055-9017
4 Maintenance
Routine maintenance of the Hi Flow
Sampler consists of the following:
Charging the battery pack
Replacing the internal filters
Replacing the gas sensors
Verifying flow rate
Setting the time and date
Resetting the microprocessor
4.1 Electrostatic Discharge Precautions
CAUTION: Electrostatic discharge (ESD) can destroy sensitive electronic components
inside the Hi Flow
Sampler when its cover is removed.
When performing any maintenance procedure that requires the Hi Flow
Sampler’s
cover to be removed, take the following precautions to prevent ESD from possibly
destroying sensitive electronic components on the main circuit board:
Make sure you have a reliable ground point available near the work site
Connect your body to the ground point using a wrist strap
Always stand on an approved conductive floor mat when touching or handling
equipment
Ground all equipment you are working on with ground straps
Handle components only on a grounded anti-static work surface
Do not wear clothing that generates static electric charges every time you move
Do not handle static generating objects while working
Maintain relative humidity in the work area between 40 and 50%
4.2 Charging the Battery Pack
WARNING: Explosion hazard! DO NOT charge the battery pack in an unsafe
atmosphere.
1. Unplug battery pack from the instrument’s power cord; then remove battery pack
from backpack.
2. Interconnect the supplied power supply, power cord, and charger as shown in
Figure 4-1. Then plug the output of the charger into the battery pack’s charger jack.
Note: Connecting the charger to the Battery Pack disables the battery’s output, making
it impossible to both charge the battery and run the instrument.
3. Allow a completely discharged battery pack to charge from 8 to 10 hours.
Vista de pagina 43